Optimizing Internal Target Margins for Moving Lung Tumors in VMAT-SBRT: A Phantom Study on Anatomical Modulation Complexity

Authors

DOI:

https://doi.org/10.63187/ampas.58

Keywords:

SBRT, VMAT, QA, Volume averaging effect, Interplay effect

Abstract

Purpose: Target motion in VMAT-based lung SBRT plans complicates dosimetric quality assurance (QA). The aim of this study is to investigate the effect of tumor locations in different anatomical lung lobes and changes in the planning target volume (PTV) on static (breath-hold) and dynamic (free-breathing) QA results.

Methodology: In this study, Monte Carlo (MC)-based VMAT SBRT plans were generated for five different lung lobes (Left Lower [LL], Left Upper [LU], Right Lower [RL], Right Middle [RM], Right Upper [RU]) using an anthropomorphic CIRS phantom. Measurements were performed both statically and dynamically (at respiratory frequencies of 0.3 Hz and 0.4 Hz and at various amplitudes) using a 2D array detector (Matrixx, 1%/1 mm) mounted on the custom-built respiratory motion simulator, as well as using an EPID (Dosisoft, 3%/1 mm). The data were analyzed using IBM SPSS Statistics v27.0.1.0 software with the Friedman and Kruskal-Wallis tests.

Findings and Conclusion: The dosimetric distortion caused by tumor motion and margin expansion depends on the target’s anatomical proximity and the complexity of MLC modulation. Our study recommends the use of a two-stage “Hybrid QA” protocol, which combines EPID flux verification and the dynamic Matrixx motion test, to distinguish between planning (dosimetric) errors and mechanical synchronization (interplay) errors.

References

1. Organization, W.H. Lung cancer. 2026; Available from: https://www.who.int/news-room/fact-sheets/detail/lung-cancer.

2. Fund, W.C.a.R. Lung Cancer. 2026; Available from: https://www.wcrf.org/preventing-cancer/cancer-types/lung-cancer/.

3. Sankar, V., R. Kothai, and N. Vanisr, Lung Cancer - A Review. International Journal of Health Sciences and Research, 2023. 13: p. 307-315.

4. Chang, J.Y., S. Senan, M.A. Paul, R.J. Mehran, A.V. Louie, P. Balter, et al., Stereotactic ablative radiotherapy versus lobectomy for operable stage I non-small-cell lung cancer: a pooled analysis of two randomised trials. The Lancet Oncology, 2015. 16(6): p. 630-637.

5. Gomez, D.R., C. Tang, J. Zhang, G.R. Blumenschein, M. Hernandez, J.J. Lee, et al., Local Consolidative Therapy Vs. Maintenance Therapy or Observation for Patients With Oligometastatic Non–Small-Cell Lung Cancer: Long-Term Results of a Multi-Institutional, Phase II, Randomized Study. Journal of Clinical Oncology, 2019. 37(18): p. 1558-1565.

6. Iyengar, P., Z. Wardak, D.E. Gerber, V. Tumati, C. Ahn, R.S. Hughes, et al., Consolidative Radiotherapy for Limited Metastatic Non–Small-Cell Lung Cancer: A Phase 2 Randomized Clinical Trial. JAMA Oncology, 2018. 4(1): p. e173501-e173501.

7. Mutsaers, A., A.V. Louie, and F. Ynoe Moraes, The case for SABR as the global standard for non-operable early-stage non-small cell lung cancer. The Lancet Regional Health – Americas, 2022. 14.

8. Palma, D.A., R. Olson, S. Harrow, S. Gaede, A.V. Louie, C. Haasbeek, et al., Stereotactic ablative radiotherapy versus standard of care palliative treatment in patients with oligometastatic cancers (SABR-COMET): a randomised, phase 2, open-label trial. The Lancet, 2019. 393(10185): p. 2051-2058.

9. Safavi, A.H., D.Y. Mak, R.G. Boldt, H. Chen, and A.V. Louie, Stereotactic ablative radiotherapy in T1-2N0M0 small cell lung cancer: A systematic review and meta-analysis. Lung Cancer, 2021. 160: p. 179-186.

10. Tandberg, D.J., B.C. Tong, B.G. Ackerson, and C.R. Kelsey, Surgery versus stereotactic body radiation therapy for stage I non-small cell lung cancer: A comprehensive review. Cancer, 2018. 124(4): p. 667-678.

11. Timmerman, R., R. McGarry, C. Yiannoutsos, L. Papiez, K. Tudor, J. DeLuca, et al., Excessive Toxicity When Treating Central Tumors in a Phase II Study of Stereotactic Body Radiation Therapy for Medically Inoperable Early-Stage Lung Cancer. Journal of Clinical Oncology, 2006. 24(30): p. 4833-4839.

12. Timmerman, R., R. Paulus, J. Galvin, J. Michalski, W. Straube, J. Bradley, et al., Stereotactic body radiation therapy for inoperable early stage lung cancer. Jama, 2010. 303(11): p. 1070-6.

13. Jiang, S.B., C. Pope, K.M. Al Jarrah, J.H. Kung, T. Bortfeld, and G.T. Chen, An experimental investigation on intra-fractional organ motion effects in lung IMRT treatments. Phys Med Biol, 2003. 48(12): p. 1773-84.

14. Court, L.E., J. Seco, X.Q. Lu, K. Ebe, C. Mayo, D. Ionascu, et al., Use of a realistic breathing lung phantom to evaluate dose delivery errors. Med Phys, 2010. 37(11): p. 5850-7.

15. Fernandez, D.J., J.T. Sick, and J.D. Fontenot, Interplay effects in highly modulated stereotactic body radiation therapy lung cases treated with volumetric modulated arc therapy. J Appl Clin Med Phys, 2020. 21(11): p. 58-69.

16. Botticella, A., A. Levy, G. Auzac, I. Chabert, C. Berthold, and C. Le Pechoux, Tumour motion management in lung cancer: a narrative review. Translational Lung Cancer Research, 2021. 10(4): p. 2011-2017.

17. Guckenberger, M., J. Wilbert, J. Meyer, K. Baier, A. Richter, and M. Flentje, Is a Single Respiratory Correlated 4D-CT Study Sufficient for Evaluation of Breathing Motion? International Journal of Radiation Oncology, Biology, Physics, 2007. 67(5): p. 1352-1359.

18. Steiner, E., C.-C. Shieh, V. Caillet, J. Booth, R. O'Brien, A. Briggs, et al., Both four-dimensional computed tomography and four-dimensional cone beam computed tomography under-predict lung target motion during radiotherapy. Radiotherapy and Oncology, 2019. 135: p. 65-73.

19. Mah, D., J. Hanley, K.E. Rosenzweig, E. Yorke, L. Braban, C.C. Ling, et al., Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. International Journal of Radiation Oncology, Biology, Physics, 2000. 48(4): p. 1175-1185.

20. Keall, P.J., G.S. Mageras, J.M. Balter, R.S. Emery, K.M. Forster, S.B. Jiang, et al., The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med Phys, 2006. 33(10): p. 3874-900.

21. Olofsson, N., K. Wikström, A. Flejmer, A. Ahnesjö, and A. Dasu, Dosimetric robustness of lung tumor photon radiotherapy evaluated from multiple event CT imaging. Phys Med, 2022. 103: p. 1-10.

22. Chetty, I.J., B. Curran, J.E. Cygler, J.J. DeMarco, G. Ezzell, B.A. Faddegon, et al., Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning. Med Phys, 2007. 34(12): p. 4818-53.

23. Ali, A.M., J.B. Greenwood, M. Varasteh, S. Esteve, P. Jeevanandam, F. Göpfert, et al., Analysis of the interplay effect in lung stereotactic ablative radiation therapy based on both breathing motion and plan characteristics. Journal of Radiotherapy in Practice, 2023. 22: p. e75.

24. Malatesta, T., A. Scaggion, F.R. Giglioli, G. Belmonte, M. Casale, P. Colleoni, et al., Patient specific quality assurance in SBRT: a systematic review of measurement-based methods. Phys Med Biol, 2023. 68(21).

25. Wang, Y., Y. Bao, L. Zhang, W. Fan, H. He, Z.W. Sun, et al., Assessment of respiration-induced motion and its impact on treatment outcome for lung cancer. Biomed Res Int, 2013. 2013: p. 872739.

26. Ekberg, L., O. Holmberg, L. Wittgren, G. Bjelkengren, and T. Landberg, What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? Radiotherapy and Oncology, 1998. 48(1): p. 71-77.

27. Erridge, S.C., Y. Seppenwoolde, S.H. Muller, M. van Herk, K. De Jaeger, J.S.A. Belderbos, et al., Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer. Radiotherapy and Oncology, 2003. 66(1): p. 75-85.

28. Grills, I.S., D. Yan, A.A. Martinez, F.A. Vicini, J.W. Wong, and L.L. Kestin, Potential for reduced toxicity and dose escalation in the treatment of inoperable non–small-cell lung cancer: A comparison of intensity-modulated radiation therapy (IMRT), 3D conformal radiation, and elective nodal irradiation. International Journal of Radiation Oncology, Biology, Physics, 2003. 57(3): p. 875-890.

29. Hanley, J., M.M. Debois, D. Mah, G.S. Mageras, A. Raben, K. Rosenzweig, et al., Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation. International Journal of Radiation Oncology, Biology, Physics, 1999. 45(3): p. 603-611.

30. Plathow, C., S. Ley, C. Fink, M. Puderbach, W. Hosch, A. Schmähl, et al., Analysis of intrathoracic tumor mobility during whole breathing cycle by dynamic MRI. International Journal of Radiation Oncology, Biology, Physics, 2004. 59(4): p. 952-959.

31. Ross, C.S., D.H. Hussey, E.C. Pennington, W. Stanford, and J. Fred Doornbos, Analysis of movement of intrathoracic neoplasms using ultrafast computerized tomography. International Journal of Radiation Oncology, Biology, Physics, 1990. 18(3): p. 671-677.

32. Seppenwoolde, Y., H. Shirato, K. Kitamura, S. Shimizu, M. van Herk, J.V. Lebesque, et al., Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. International Journal of Radiation Oncology, Biology, Physics, 2002. 53(4): p. 822-834.

33. Shimizu, S., H. Shirato, S. Ogura, H. Akita-Dosaka, K. Kitamura, T. Nishioka, et al., Detection of lung tumor movement in real-time tumor-tracking radiotherapy. International Journal of Radiation Oncology, Biology, Physics, 2001. 51(2): p. 304-310.

34. Sixel, K.E., M. Ruschin, R. Tirona, and P.C.F. Cheung, Digital fluoroscopy to quantify lung tumor motion: potential for patient-specific planning target volumes. International Journal of Radiation Oncology, Biology, Physics, 2003. 57(3): p. 717-723.

35. Keall, P.J., G.S. Mageras, J.M. Balter, R.S. Emery, K.M. Forster, S.B. Jiang, et al., The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Medical Physics, 2006. 33(10): p. 3874-3900.

36. Videtic, G.M., C. Hu, A.K. Singh, J.Y. Chang, W. Parker, K.R. Olivier, et al., A Randomized Phase 2 Study Comparing 2 Stereotactic Body Radiation Therapy Schedules for Medically Inoperable Patients With Stage I Peripheral Non-Small Cell Lung Cancer: NRG Oncology RTOG 0915 (NCCTG N0927). Int J Radiat Oncol Biol Phys, 2015. 93(4): p. 757-64.

37. Low, D.A., J.M. Moran, J.F. Dempsey, L. Dong, and M. Oldham, Dosimetry tools and techniques for IMRT. Med Phys, 2011. 38(3): p. 1313-38.

38. van Elmpt, W., L. McDermott, S. Nijsten, M. Wendling, P. Lambin, and B. Mijnheer, A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol, 2008. 88(3): p. 289-309.

39. Boggula, R., M. Birkner, F. Lohr, V. Steil, F. Wenz, and H. Wertz, Evaluation of a 2D detector array for patient-specific VMAT QA with different setups. Phys Med Biol, 2011. 56(22): p. 7163-77.

40. Woon, W., P.B. Ravindran, P. Ekayanake, V. S, Y.Y. Lim, and J. Khalid, A study on the effect of detector resolution on gamma index passing rate for VMAT and IMRT QA. J Appl Clin Med Phys, 2018. 19(2): p. 230-248.

41. Rao, M., J. Wu, D. Cao, T. Wong, V. Mehta, D. Shepard, et al., Dosimetric impact of breathing motion in lung stereotactic body radiotherapy treatment using intensity modulated radiotherapy and volumetric modulated arc therapy [corrected]. Int J Radiat Oncol Biol Phys, 2012. 83(2): p. e251-6.

42. Ong, C., W.F. Verbakel, J.P. Cuijpers, B.J. Slotman, and S. Senan, Dosimetric impact of interplay effect on RapidArc lung stereotactic treatment delivery. Int J Radiat Oncol Biol Phys, 2011. 79(1): p. 305-11.

43. Edvardsson, A., F. Nordström, C. Ceberg, and S. Ceberg, Motion induced interplay effects for VMAT radiotherapy. Phys Med Biol, 2018. 63(8): p. 085012.

44. Poppe, B., A. Blechschmidt, A. Djouguela, R. Kollhoff, A. Rubach, K.C. Willborn, et al., Two-dimensional ionization chamber arrays for IMRT plan verification. Med Phys, 2006. 33(4): p. 1005-15.

45. Masi, L., R. Doro, V. Favuzza, S. Cipressi, and L. Livi, Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys, 2013. 40(7): p. 071718.

Downloads

Published

2026-09-19

How to Cite

YÜREKLİ, A. F., & DİRİCAN, B. (2026). Optimizing Internal Target Margins for Moving Lung Tumors in VMAT-SBRT: A Phantom Study on Anatomical Modulation Complexity. Advances in Medical Pyhsics and Applied Sciences, 2(3), 102–111. https://doi.org/10.63187/ampas.58

Similar Articles

1 2 > >> 